超低剂量光子计数CT用于基于解剖的人工耳蜗植入适配的可行性:一项尸体概念验证研究。
Feasibility of Ultra-Low-Dose Photon-Counting CT for Anatomy-Based Cochlear Implant Fitting: A Cadaveric Proof-of-Concept Study.
文献信息
| PMID | 43070293 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Thore Schade-Mann |
| 作者单位 | Department of Otolaryngology-Head & Neck Surgery, University of Tübingen Medical Center, Tübingen, Germany. |
| 期刊 | Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology |
| SCI 分区 | Q2 |
| IF | 1.9 |
| 研究类型 | 临床研究 · 临床 |
| 所属专科 | 耳科 |
中文摘要
假设: 在术后光子计数探测器CT(PCD-CT)成像过程中逐步降低CT管电流,可在低至临床标准剂量5%的剂量下保持基于解剖的人工耳蜗植入适配所需的频率估计准确性。
背景: 基于解剖的适配(ABF)利用术后CT将电极触点分配到耳蜗频率图,减少频率-位置失配,并可能改善人工耳蜗(CI)结局。虽然PCD-CT较传统CT降低了辐射,但兼容ABF的最低术后剂量尚未确定。
方法: 一具人体供体在经圆窗途径插入CI测试电极之前和之后,以100%、50%、25%、10%和5%管电流接受PCD-CT(NAEOTOM Alpha,Siemens Healthineers)成像。使用OTOPLAN软件(版本4,Cascination)提取电极角插入深度和频率位置。以100%参考扫描为基准,以半音为单位量化频率偏差。分析还使用50%术前扫描作为替代参考标准重复进行,扩展了先前发表的术前剂量降低研究结果。
结果: 对于所评估的耳蜗内触点,在所有剂量水平下电极检测均成功。使用100%术前参考时,5%术后剂量下的平均频率偏差为0.02±0.29半音,所评估触点均未超过一个半音。当以50%术前扫描作为参考时,术后剂量≥10%时偏差仍≤1半音;在5%术后剂量下,三个相邻触点略微超过该阈值(最大:-1.20半音)。
结论: 在这项单供体概念验证研究中,PCD-CT在机构参考剂量的5%下仍保持电极可见性,并对所评估的耳蜗内触点产生较小的频率估计偏差。因此,这些发现支持在所评估电极区域内实现大幅术后剂量降低的技术可行性。需要在更大临床队列中使用完整插入进行验证。
英文摘要
HYPOTHESIS: Stepwise reduction of CT tube current during postoperative photon-counting detector CT (PCD-CT) imaging preserves the frequency-estimation accuracy required for anatomy-based cochlear-implant fitting at doses as low as 5% of the clinical standard.
BACKGROUND: Anatomy-based fitting (ABF) uses postoperative CT to allocate electrode contacts to the cochlear frequency map, reducing frequency-to-place mismatch and potentially improving cochlear implant (CI) outcomes. While PCD-CT lowers radiation compared with conventional CT, the minimum postoperative dose compatible with ABF has not been established.
METHODS: A human body donor was imaged with PCD-CT (NAEOTOM Alpha, Siemens Healthineers) before and after CI test electrode insertion via a round window approach at 100%, 50%, 25%, 10%, and 5% tube current. Electrode angular insertion depth and frequency position were extracted using OTOPLAN software (Version 4, Cascination). Frequency deviations from the 100% reference were quantified in semitones. Analysis was repeated using the 50% preoperative scan as an alternative reference standard, extending previously published preoperative dose-reduction findings.
RESULTS: Electrode detection was successful at all dose levels for the evaluated intracochlear contacts. Using the 100% preoperative reference, mean frequency deviation at 5% postoperative dose was 0.02±0.29 semitones, with no evaluated contact exceeding one semitone. When the 50% preoperative scan served as reference, deviations remained ≤1 semitone for postoperative doses ≥10%; three adjacent contacts marginally exceeded this threshold at 5% postoperative dose (maximum: -1.20 semitones).
CONCLUSIONS: In this single-donor proof-of-concept study, PCD-CT at 5% of the institutional reference dose retained electrode visibility and yielded small frequency-estimation deviations for the evaluated intracochlear contacts. These findings therefore support the technical feasibility of substantial postoperative dose reduction within the evaluated electrode region. Validation using complete insertions in larger clinical cohorts is required.