两种不同系统的主动骨传导植入处理器可以互换吗?一项技术与临床可行性研究。

Can active bone conduction implant processors of two different systems be interchanged? A technical and clinical feasibility study.

临床研究耳科IF 3.4Q1

文献信息

中文摘要

目的: 主动经皮骨传导植入系统被设计为封闭的硬件系统,其中外部音频处理器仅与其对应的植入式换能器配合工作。然而,不同制造商组件的机电兼容性此前尚未被研究。本研究评估了使用Bonebridge(MED-EL)和Sentio(Oticon Medical)系统的跨系统处理器-植入物组合的技术可行性、输出性能和初步临床适用性。
方法: 将Bonebridge BCI 602和Sentio Ti植入物与颅骨模拟器和植入物模型耦合。在标准化参考测试设置下评估Samba 2 BB和Sentio 1 Mini音频处理器的原生和非原生组合。使用频率响应和最大输出记录测量力输出特性。此外,两名植入患者接受了替代处理器系统的急性实验性验配。听力学评估包括助听听场阈值、安静和噪声下的言语识别、分类响度缩放以及言语、空间和听觉质量量表(SSQ12)。
结果: 在所有研究的处理器-植入物组合中,跨系统激活在技术上均可行。颅骨模拟器测量显示原生与非原生配对之间的输出力水平存在显著差异,4PTA力输出差异高达21.6 dB,尽管反馈相关干扰影响了高输入水平下的测量。模型测量确认所有组合均可听。在两项临床评估中,非原生处理器配置均实现了成功验配和开放式言语理解。言语识别性能、助听阈值和主观听力结果在原生与非原生配置之间总体相当,两名参与者均报告在至少部分结局指标中偏好Sentio 1 Mini处理器。
结论: 所测试的音频处理器在技术上能够激活非原生植入物,并且在两项监督下的临床评估中短期言语感知是可能的。然而,输出特性的显著差异表明系统特异性的机电优化。但是,输出特性的差异以及缺乏系统性的电学、热学和长期安全性评估意味着不能推荐临床互换性。在考虑跨系统使用之前,需要进一步的技术和临床评估。

英文摘要

OBJECTIVES: Active transcutaneous bone conduction implant systems are designed as closed hardware systems in which external audio processors operate exclusively with their corresponding implanted transducers. However, the electro-mechanical compatibility of components from different manufacturers has not previously been investigated. This study evaluated the technical feasibility, output performance, and preliminary clinical applicability of cross-system processor-implant combinations using the Bonebridge (MED-EL) and Sentio (Oticon Medical) systems.
METHODS: A Bonebridge BCI 602 and a Sentio Ti implant were coupled to a skull simulator and implant phantoms. Native and non-native combinations of the Samba 2 BB and Sentio 1 Mini audio processors were evaluated under standardized reference test settings. Force output characteristics were measured using frequency response and maximum output recordings. In addition, two implanted patients underwent acute experimental fitting with the alternative processor system. Audiological assessment included aided sound-field thresholds, speech recognition in quiet and noise, categorical loudness scaling, and the Speech, Spatial and Qualities of Hearing Scale (SSQ12).
RESULTS: Cross-system activation was technically possible in all investigated processor-implant combinations. Skull simulator measurements demonstrated substantial differences in output force levels between native and non-native pairings, with 4PTA force-output differences of up to 21.6 dB, although feedback-related interference affected measurements at high input levels. Phantom measurements confirmed audibility for all combinations. In both clinical assessments, successful fitting and open-set speech understanding were achieved with the non-native processor configuration. Speech recognition performance, aided thresholds, and subjective hearing outcomes were generally comparable between native and non-native configurations, with both participants reporting a preference for the Sentio 1 Mini processor in at least some outcome measures.
CONCLUSIONS: The tested audio processors were technically capable of activating non-native implants, and short-term speech perception was possible in two supervised clinical assessments. Nevertheless, substantial differences in output characteristics indicate system-specific electro-mechanical optimization. However, differences in output characteristics and the lack of systematic electrical, thermal, and long-term safety evaluations mean that clinical interchangeability cannot be recommended. Further technical and clinical evaluation is required before cross-system use can be considered.