将电诱发复合动作电位阈值与基于验证的计算机断层扫描电极定位的人工耳蜗位置相关联
Correlating Electrically Evoked Compound Action Potential Thresholds with Cochlear Implant Position Using Validated Computed Tomography-Based Electrode Localization.
文献信息
| PMID | 43061776 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Fergio Sismono |
| 作者单位 | European Institute for ORL-HNS, ZAS Hospitals Antwerp, Wilrijk, Antwerp, Belgium. |
| 期刊 | The journal of international advanced otology |
| SCI 分区 | Q3 |
| IF | 1.3 |
| 研究类型 | 临床研究 · 临床 |
| 所属专科 | 耳科 |
中文摘要
背景: 电诱发复合动作电位阈值(TECAP)是快速、可及且客观的测量方法,用于评估人工耳蜗植入者的电极-神经元界面。虽然通常假设TECAP水平与电极位置共变,但既往电极定位方法的敏感性有限,迄今使对该关系的详细分析变得复杂。本研究利用高分辨率锥形束计算机断层扫描(CT)和经过验证的算法,重新评估电极位置与阈值轮廓之间的关系。
方法: 回顾性分析16名植入Cochlear Nucleus Contour Advance阵列的成年受者的术中TECAP和CT扫描。确定电极至蜗轴壁距离(EMD)、电极至Rosenthal管距离(ERD)和中蜗轴轴距离,并与阈值轮廓进行相关性分析。
结果: 基于每个受者的线性回归分析证实,TECAP随EMD和ERD显著增加。在16名受者中的13名中,电极至Rosenthal管距离产生的R2值高于EMD,并在3名个体中表现出近乎完美的相关性(R2 > 0.8,P < .001)。回归斜率和截距也观察到显著的受者间变异性,这可能部分反映受者特异性神经健康。拟合了考虑这种变异性的线性混合模型,解释了总方差的69.4%(R2 > 0.694,P < .001)。
结论: 本研究结果表明,TECAP轮廓与内外侧电极位置相关,并可能受受者特异性神经健康差异的影响。将来自专门CT分析算法的电极位置纳入神经健康测量,可能有助于分离神经方面并实现新的治疗应用。
英文摘要
BACKGROUND: Electrically evoked compound action potential thresholds (TECAP) are quick, accessible, and objective measurements used to evaluate the electrode-to-neuron interface in cochlear implant recipients. While TECAP levels are assumed to covary with electrode position, limited sensitivity of previous electrode localization methods has thus far complicated a detailed analysis of this relationship. This study utilized high-resolution cone-beam computed tomography (CT) and a validated algorithm in order to reevaluate the relationship between electrode position and threshold profiles.
METHODS: Intra-operative TECAPs and CT scans from 16 adult recipients of Cochlear Nucleus Contour Advance arrays were retrospectively analyzed. Electrode-to-modiolar wall distances (EMDs), electrode-to-Rosenthal's canal distances (ERDs) and mid-modiolar axis distances were determined and correlated with the threshold profiles.
RESULTS: Linear regression analysis on a per subject basis confirmed that TECAP significantly increased with EMD and ERD. Electrode-toRosenthal's canal distances yielded higher R2 values than EMD in 13 of 16 subjects and demonstrated near-perfect correlation in 3 individuals (R2 > 0.8, P < .001). Substantial inter-subject variability was also observed in regression slopes and intercepts, which may partly reflect subject-specific neural health. A linear mixed model accounting for this variability was fitted and explained 69.4% of the total variance (R2> 0.694, P < .001).
CONCLUSION: The results from this study suggest that TECAP profiles are associated with mediolateral electrode position and are possibly affected by subject-specific differences in neural health. Incorporating electrode position from specialized CT analysis algorithms into neural health measures may help isolate the neural aspects and enable novel therapeutic applications.