使用光学相干断层扫描的耳蜗植入手术高分辨率术中导航
High-Resolution Intraoperative Guide for Cochlear Implant Surgery Using Optical Coherence Tomography.
文献信息
| PMID | 43052069 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Guillaume Gersdorff |
| 作者单位 | Department of Surgery (Otolaryngology), University of Melbourne, East Melbourne, VIC, Australia. |
| 期刊 | Audiology & neuro-otology |
| SCI 分区 | Q3 |
| IF | 1.8 |
| 研究类型 | 基础研究 · 基础/转化 |
| 所属专科 | 耳科 |
中文摘要
引言: 尽管软手术技术有所进步,耳蜗植入(CI)电极插入过程中的耳蜗内创伤仍会发生,可能损伤耳蜗微结构并影响听力结果或未来治疗。虽然圆窗膜(RWM)入路通常创伤较小,但钩区的复杂解剖结构仍可导致机械性损伤,通常是由于视觉引导有限。术前影像和术中光学相干断层扫描(OCT)可通过提供耳蜗微结构的高分辨率实时可视化来减少创伤。
方法: 纳入8块新鲜冷冻人类颞骨(TB1-TB8)。前6个标本(TB1-TB6)接受显微CT扫描并手动分割以建立详细解剖基线,并用于OCT-显微CT共配准。按照标准耳蜗植入入路,使用显微镜集成OCT系统可视化RWM后的耳蜗内结构,并通过前下耳蜗开窗进行额外成像以比较入路。在选定标本中于OCT引导下进行电极插入,同时在TB7-TB8中于透视控制下评估术中工作流程和插入动力学的可行性。
结果: 每个颞骨均成功获得OCT图像,适当的RWM暴露至关重要。OCT显微镜能够在手术工作距离下从各种轴、深度和角度扫描RWM和鼓阶(ST),改善近端ST解剖的评估。OCT似乎可可靠地通过RWM评估微解剖,指导CI电极选择、入路和插入过程中近端阵列路径的修改。通过可视化关键结构并推断更深部结构,它支持基于术前影像的规划,并可能降低耳蜗内损伤的风险。
结论: 术中OCT成像提高了耳蜗开口部位和电极插入轴的精确度。通过整合OCT,CI手术可能从很大程度上盲目的手术转变为具有显著视觉引导的手术,减少手术创伤并改善术后听力保留。
英文摘要
INTRODUCTION: Despite advances in soft surgery, intracochlear trauma during cochlear implant (CI) electrode insertion still occurs, potentially damaging cochlear microstructures and affecting hearing outcomes or future treatments. While the round window membrane (RWM) approach is generally less traumatic, the hook region's complex anatomy can still cause mechanical injury, often due to limited visual guidance. Preoperative imaging and intraoperative optical coherence tomography (OCT) could reduce trauma by offering high-resolution, real-time visualization of cochlear microstructures.
METHODS: Eight fresh-frozen human temporal bones (TB1-TB8) were included. The first 6 specimens (TB1-TB6) underwent micro-CT scanning with manual segmentation to establish detailed anatomical baselines and served for OCT-micro-CT co-registration. Following a standard cochlear implantation approach, intracochlear structures behind the RWM were visualized using a microscope-integrated OCT system, with additional imaging performed via an anteroinferior cochleostomy to compare access routes. Electrode insertions were performed under OCT guidance in selected specimens, while feasibility of intraoperative workflow and insertion dynamics was assessed in TB7-TB8 under fluoroscopic control.
RESULTS: OCT images were successfully obtained for each TB, with proper RWM exposure being crucial. The OCT microscope enabled scanning of the RWM and scala tympani (ST) from various axes, depths, and angles at a surgical working distance, improving assessment of proximal ST anatomy. OCT appears reliable for evaluating microanatomy through the RWM, guiding CI electrode selection, approach, and proximal array course modification during insertion. By visualizing critical structures and inferring deeper ones, it supports preoperative imagery-based planning and may reduce the risk of intracochlear damage.
CONCLUSION: Intraoperative OCT imaging enhances the precision of the cochlear opening site and electrode insertion axis. By integrating OCT, CI surgery could shift from a largely blind procedure to one with significant visual guidance, reducing surgical trauma and improving postoperative hearing preservation.