坡度型高频听力损失CI632植入者的听力保留与疗效结局:多中心试验结果
Hearing Preservation and Efficacy Outcomes in CI632 Recipients with Sloping High-Frequency Hearing Loss: Multicenter Trial Results.
文献信息
| PMID | 43052065 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Sarah Mowry |
| 作者单位 | Department of Otolaryngology-Head and Neck Surgery, University Hospitals Cleveland Medical Center, Cleveland, OH, USA. |
| 期刊 | Audiology & neuro-otology |
| SCI 分区 | Q3 |
| IF | 1.8 |
| 研究类型 | 临床研究 · 临床 |
| 所属专科 | 耳科 |
中文摘要
引言: 具有显著低频残余听力的成人越来越多地被考虑接受人工耳蜗植入。本研究评估了在扩大适应证下使用细长近蜗轴电极(CI632)的成人听力保留和言语表现结局。
方法: 在一项于美国三家三级中心开展的前瞻性、开放标签、单臂可行性试验中,十名成人植入耳500 Hz低频阈值≤50 dB HL、呈坡度型高频损失且助听后CNC评分≤60%,接受了CI632装置。参与者开机后随访12个月。评估包括气导和骨导测听、电极阻抗、单侧和双侧安静下CNC及噪声下AzBio句子测试、患者报告结局以及术后计算机断层扫描(CT)成像。
结果: 低频纯音平均听阈(125-500 Hz)在开机时平均偏移19.3 dB,并在12个月时保持稳定于22.4 dB。各次访视阻抗稳定。由于保留了声学听力,所有参与者均在植入耳使用电声刺激。所有言语识别指标均显著改善,12个月时平均改善幅度依测试条件不同为22至54个百分点。患者报告结局亦有改善。CT成像证实所有电极均位于鼓阶,平均角度插入深度和包绕因子分别为406.6°和59.4%。
结论: 细长近蜗轴电极可在扩大适应证下植入的成人中保留低频声学听力,并在客观和主观结局上均观察到有意义的改善。该队列的听力保留结局超过了近蜗轴阵列通常报告的结局。尽管所有参与者均实现了功能上可助听的听力保留,但样本量小限制了这一发现的普遍性,需要更大规模的研究来证实这些结果。
英文摘要
INTRODUCTION: Adults with substantial residual low-frequency (LF) hearing are increasingly considered for cochlear implantation. This study evaluated hearing preservation and speech performance outcomes using a slim perimodiolar electrode (CI632) in adults under expanded indications.
METHODS: In a prospective, open-label, single-arm feasibility trial at three US tertiary centers, ten adults with LF thresholds ≤50 dB HL at 500 Hz, sloping high-frequency loss, and aided CNC scores ≤60% in the ear to be implanted received the CI632 device. Participants were followed for 12 months postactivation. Assessments included air and bone conduction audiometry, electrode impedances, unilateral and bilateral CNC in quiet and AzBio sentences in noise, patient-reported outcomes, and postoperative computerized tomography (CT) imaging.
RESULTS: The LF pure-tone average (125-500 Hz) showed a mean shift of 19.3 dB at activation and remained stable at 22.4 dB at 12 months. Impedances were stable across visits. Due to preserved acoustic hearing, all participants utilized electric-acoustic stimulation in the implant ear. Speech recognition improved significantly across all measures, with an average improvement at 12 months ranging from 22 to 54 percentage points, depending on the condition. Patient-reported outcomes also improved. CT imaging confirmed scala tympani placement for all electrodes, with a mean angular insertion depth and wrapping factor of 406.6° and 59.4%, respectively.
CONCLUSION: The slim perimodiolar electrode can preserve LF acoustic hearing in adults implanted under expanded indications, with meaningful improvements observed in both objective and subjective outcomes. Hearing preservation outcomes in this cohort exceeded those typically reported for perimodiolar arrays. While all participants achieved functionally aidable hearing preservation, the small sample size limits the generalizability of this finding, and larger studies are needed to confirm these results.