振动人工耳蜗阵列用于声刺激的效率
Efficiency of a vibrating cochlear implant array for acoustic stimulation.
文献信息
| PMID | 43038721 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Mohammad Ghoncheh |
| 作者单位 | Department of Otolaryngology and Institute of Audioneurotechnology (VIANNA), Hannover Medical School, Hannover, Germany; Cluster of Excellence "Hearing4all", Hannover Medical School, Hannover, Germany. Electronic address: Ghoncheh.Mohammad@MH-Hannover.de. |
| 期刊 | Hearing research |
| SCI 分区 | Q1 |
| IF | 3.2 |
| 研究类型 | 基础研究 · 基础/转化 |
| 所属专科 | 耳科 |
中文摘要
引言: 患有传导性或混合性听力损失的患者可从声刺激或机械刺激与人工耳蜗(CI)的结合中获益。我们在人颞骨(TB)中实验性研究了将CI电极阵列用作声-机械刺激器的可行性。
方法: 首先,在颞骨标本(N = 10)中测量镫骨对外耳道声刺激的位移幅度。测量声刺激引起的前庭阶(SV)耳蜗内压(PSV)。随后,将CI(FLEX20,MED-EL,奥地利)电极经圆窗插入鼓阶(ST),并使用定制适配器将其远端连接到商用压电换能器。最后,测量对插入的CI电极进行机械刺激所引起的电极位移和PSV。确定电极位移1 µm时的等效声压级(eq. dB SPL),以评估振动CI所达到的输出水平。
结果: 在低频(125 - 500 Hz)下,振动CI在电极位移1 µm时产生的平均等效输出水平为108 eq. dB SPL(LDisp)。在1 kHz以上,输出每倍频程增加约10 dB,达到103-139 eq. dB SPL,在高达10 kHz的各实验和频率间变异性为8-15 dB(SD)。
结论: 研究结果表明,CI电极1 µm的振动可提供约110 eq. dB SPL的平均输出,可能与EAS所需的输出相容。
英文摘要
INTRODUCTION: Patients with conductive or mixed hearing loss can benefit from a combination of acoustic or mechanical stimulation and a cochlear implant (CI). We experimentally investigated the feasibility of using the CI electrode array as an acousto-mechanical stimulator in human temporal bones (TBs).
METHODS: First, the displacement magnitude of the stapes in response to the acoustic stimulation in the external auditory canal was measured in temporal bone specimens (N = 10). The scala vestibuli (SV) intracochlear pressure (PSV) in response to the acoustic stimulation was measured. Next, a CI (FLEX20, MED-EL, Austria) electrode was inserted into the scala tympani (ST) through the round window, and its distal end was attached to a commercial piezoelectric transducer using a custom-made adapter. Finally, the electrode displacement and PSV was measured in response to mechanical stimulation of the inserted CI electrode. The equivalent sound pressure level (eq. dB SPL) at 1 µm electrode displacement was determined to evaluate the achieved output level of the vibrating CI.
RESULTS: The vibrating CI produced an average equivalent output level of 108 eq. dB SPL (LDisp) at low frequencies (125 - 500 Hz) for a 1 µm electrode displacement of the CI. Above 1 kHz, the output increased by ∼10 dB per octave, reaching 103-139 eq. dB SPL, with a variability of 8-15 dB (SD) across experiments and frequencies up to 10 kHz.
CONCLUSION: The findings indicate that a 1 µm vibration of a CI electrode can provide an average output of ∼110 eq. dB SPL, potentially compatible with the output required for EAS.