人工耳蜗植入术中应用4K3D外视镜对面神经隐窝进行术中非接触式三维测量
Intraoperative non-contact three-dimensional measurement of the facial recess using a 4K3D exoscope during cochlear implantation.
文献信息
| PMID | 43037709 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Chaoyue Zhao |
| 作者单位 | Nankai University, Tianjin, 300071, China. |
| 期刊 | European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery |
| SCI 分区 | Q1 |
| IF | 2.3 |
| 研究类型 | 临床研究 · 临床 |
| 所属专科 | 耳科 |
中文摘要
目的: 评估基于4K3D外视镜的非接触式三维面神经隐窝测量技术的可行性,并在尸体标本中探讨其与经比例校准的二维ImageJ测量的一致性,以及在接受人工耳蜗植入的患者中与术前CT测量的一致性。
方法: 本研究使用的系统将AINNOVI-100外视镜与SurgeReal 3D Studio相结合。尸体可行性评估和方法比较纳入来自8具福尔马林固定尸体头部的10块颞骨。将三维测量结果与经比例校准的二维ImageJ测量结果进行比较。前瞻性单中心队列纳入23例接受人工耳蜗植入的患者。采用绝对一致性组内相关系数和Bland-Altman分析,将术中测量结果与术前高分辨率计算机断层扫描测量结果进行比较,以进行探索性一致性分析。
结果: 尸体标本的组内相关系数范围为0.878至0.934。平均模型生成时间为1.97秒,从重建开始到结果存储的记录工作流程持续46.07秒。35次尸体测量尝试中31次成功(88.6%),包括29次首次成功(82.9%);4次尝试失败(11.4%)。在23例临床病例中,面神经隐窝宽度的一致性最强(组内相关系数,0.962)。角度的一致性界限较宽,但在两个队列中均保持在所述的±10°比较范围内。
结论: 本研究支持使用4K3D外视镜对暴露的面神经隐窝进行非接触式定量测量的可行性。观察到的一致性,尤其是面神经隐窝宽度的一致性,以及较短的重建和测量时间,支持进一步评估该工作流程用于术中解剖记录。未来研究应确立三维测量的准确性和可重复性,并探讨所得信息如何有助于手术决策。
英文摘要
PURPOSE: To evaluate the technical feasibility of 4K3D exoscope-based non-contact three-dimensional measurement of the facial recess and explore agreement with scale-calibrated two-dimensional ImageJ measurements in cadaveric specimens and preoperative CT measurements in patients undergoing cochlear implantation.
METHODS: The system used in this study combined an AINNOVI-100 exoscope with SurgeReal 3D Studio. Cadaveric feasibility assessment and method comparison included 10 temporal bones from eight formalin-fixed cadaveric heads. Three-dimensional measurements were compared with scale-calibrated two-dimensional ImageJ measurements. The prospective single-center cohort included 23 patients undergoing cochlear implantation. Intraoperative measurements were compared with preoperative high-resolution computed tomography measurements for exploratory agreement using absolute-agreement intraclass correlation coefficients and Bland-Altman analysis.
RESULTS: Cadaveric intraclass correlation coefficients ranged from 0.878 to 0.934. Mean model-generation time was 1.97 s, and the recorded workflow from reconstruction initiation to result storage lasted 46.07 s. 31 of 35 cadaveric measurement attempts were successful (88.6%), including 29 initial successes (82.9%); 4 attempts failed (11.4%). In the 23 clinical cases, agreement was strongest for facial recess width (intraclass correlation coefficient, 0.962). Angular limits of agreement were wider but remained within the stated ± 10° comparison margin in both cohorts.
CONCLUSION: This study supports the feasibility of obtaining non-contact quantitative measurements of the exposed facial recess using a 4K3D exoscope. The observed agreement, particularly for facial recess width, and the short reconstruction and measurement times support further evaluation of this workflow for intraoperative anatomical documentation. Future studies should establish three-dimensional measurement accuracy and reproducibility and examine how the resulting information contributes to surgical decision-making.