人工与机器人电极插入后人工耳蜗植入听力结果的比较
Comparison of Cochlear Implant Hearing Outcomes Following Manual and Robotic Electrode Insertion.
文献信息
| PMID | 43037006 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Ilana Yellin |
| 作者单位 | Department of Otolaryngology-Head and Neck Surgery, AdventHealth Celebration, Celebration, Florida, USA. |
| 期刊 | Otolaryngology--head and neck surgery : official journal of American Academy of Otolaryngology-Head and Neck Surgery |
| SCI 分区 | Q1 |
| IF | 2.9 |
| 研究类型 | 临床研究 · 临床 |
| 所属专科 | 耳科 |
中文摘要
目的: 本研究旨在比较接受机器人辅助人工耳蜗植入(CI)的患者与接受手动CI的患者的初始低频听力保留(LFHP)。
研究设计: 回顾性综述。
地点: 三级转诊中心。
方法: 本研究纳入具有低频残余听力的成人CI患者,低频残余听力定义为术前250和500 Hz低频纯音平均听阈(LFPTA)≤80 dB HL。患者接受机器人辅助或手动CI。所有植入体均使用侧壁电极。主要结局为机器人辅助或手动CI后LFPTA变化(ΔLFPTA)及LFHP的存在,LFHP定义为ΔLFPTA≤30 dB HL。
结果: 35例机器人辅助和27例手动病例符合纳入标准,用于比较初始LFHP。机器人辅助插入和手动插入的术前平均LFPTA分别为56.14和62.04 dB HL(P=.11),术后平均LFPTA分别为88.86和91.02 dB HL(P=.63)。机器人辅助插入组与手动插入组之间的ΔLFPTA无显著差异(机器人辅助32.17 dB HL vs 手动28.98 dB HL,P=.82)。两组间LFHP相似(机器人辅助57.10% vs 手动51.90%,P=.68)。
结论: 在本研究组中,机器人辅助与手动CI在初始LFHP方面无显著差异,但机器人辅助插入可能仍对随时间推移保留残余听力有益。需要更长期随访的进一步研究来充分评估机器人辅助插入对听力保留的影响。
英文摘要
OBJECTIVE: The purpose of this study is to compare initial low frequency hearing preservation (LFHP) in patients undergoing robotic-assisted cochlear implantation (CI) to those undergoing manual CI.
STUDY DESIGN: Retrospective review.
SETTING: Tertiary referral center.
METHODS: Adult CI patients with low-frequency residual hearing, defined as a preoperative low-frequency pure tone average (LFPTA) at 250 and 500 Hz ≤80 dB HL, were included in this study. Patients underwent either robotic-assisted or manual CI. All implants utilized lateral wall electrodes. The primary outcomes were the change in LFPTA (ΔLFPTA) and the presence of LFHP, defined as ΔLFPTA ≤ 30 dB HL, following robotic-assisted or manual CI.
RESULTS: 35 robotic-assisted and 27 manual cases met the inclusion criteria for comparison of initial LFHP. The mean preoperative LFPTA was 56.14 and 62.04 dB HL (P = .11) while the mean postoperative LFPTA was 88.86 and 91.02 dB HL for robotic-assisted and manual insertion, respectively (P = .63). The ΔLFPTA was not significantly different between the robotic-assisted and manual insertion groups (32.17 dB HL robotic-assisted vs 28.98 dB HL manual, P = .82). LFHP was similar between groups (57.10% robotic-assisted vs 51.90% manual, P = .68).
CONCLUSION: There is no significant difference in initial LFHP between robotic-assisted and manual CI in this study group, though robotic-assisted insertion may still offer benefit for preservation of residual hearing over time. Further studies with longer follow-up are needed to fully evaluate the impact of robotic-assisted insertion on hearing preservation.