光子计数CT用于人工耳蜗术后成像:与平板探测器CT的比较

Photon-Counting CT for Postoperative Imaging of Cochlear Implants: Comparison to Flat-Detector CT.

临床研究耳科IF 6.6Q1

文献信息

中文摘要

目的: 自2021年获FDA批准以来,光子计数探测器计算机断层扫描(PCCT)已证明其在高分辨率成像中的可行性,同时可降低辐射剂量。高分辨率显示人工耳蜗(CI)的完整性和位置对于术后管理至关重要。本研究旨在评估PCCT在CI植入术后成像中的表现,并与当前临床金标准——高分辨率平板探测器CT(FDCT)进行比较。
材料与方法: 回顾性分析2024年8月至2026年4月期间CI植入术后的PCCT扫描。若在同一患者序贯双侧人工耳蜗植入过程中,FDCT采集先于PCCT,则纳入分析。由2名经验丰富的阅片者以共识方式使用评分系统评估图像质量,分析内耳和中耳的结构显示、CI完整性、金属伪影和运动伪影,并计算总体评分。辐射剂量从图像归档和通信系统中获取。
结果: 共评估28例CI植入术后患者的PCCT扫描,FDCT与PCCT扫描之间的平均时间间隔为4.5±5.1年。PCCT的图像质量在所有单项成像特征上均不劣于FDCT(P<0.001,例如单个CI电极的区分:中位数[四分位距] PCCT 3[3至3];FDCT 3[3至3];P<0.001),但骨结构均匀性和乳突气房分隔除外(P=0.999)。PCCT在减少金属伪影和运动伪影以及提高镫骨显示方面优于FDCT(P=0.009至<0.001;例如金属伪影PCCT 3[2.75至3];FDCT 2[1至2];P<0.001)。采用本机构方案时,PCCT的有效辐射剂量为0.05±0.01 mSv,比FDCT的0.44±0.29 mSv低8.8±5.4倍(P<0.001)。
结论: 就CI术后评估的图像质量而言,PCCT不劣于FDCT,或在单项特征上优于FDCT,同时辐射剂量显著更低、扫描时间更短。

英文摘要

OBJECTIVES: After FDA approval in 2021, photon-counting detector computed tomography (PCCT) has demonstrated its feasibility for high-resolution imaging while lowering the radiation dose at the same time. High-resolution visualization of cochlear implant (CI) integrity and positioning is crucial for postoperative management. The aim of this study was to evaluate the performance of PCCT for postoperative imaging after CI implantation in comparison to high-resolution flat-detector CT (FDCT), the current clinical gold standard.
MATERIALS AND METHODS: Postoperative PCCT scans after cochlear implantation were retrospectively analyzed from August 2024 to April 2026. Scans were included for analysis if FDCT acquisition had been before PCCT on the same patient in the course of sequential bilateral cochlear implantation. Image quality was assessed in consensus by 2 experienced readers using a scoring system analyzing structural delineation of the inner and middle ear, CI integrity, metal and motion artifacts as well as calculating an overall score. Radiation dose was retrieved from the picture archiving and communication system.
RESULTS: Twenty-eight postoperative PCCT scans of patients after CI implantation with a mean time delay of 4.5±5.1 years between FDCT and PCCT scans were evaluated. Image quality of PCCT was noninferior to FDCT for all single imaging features (P<0.001, eg, distinction of single CI electrodes: median [interquartile range] PCCT 3 [3 to 3]; FDCT 3 [3 to 3]; P<0.001) with the exception of the homogeneity of bony structures and the mastoid cell septation (P=0.999). PCCT was superior to FDCT in reducing metal and motion artifacts and increasing the delineation of the stapes (P=0.009 to <0.001; eg, metal artifacts PCCT 3 [2.75 to 3]; FDCT 2 [1 to 2]; P<0.001). Effective radiation dose of PCCT was 0.05±0.01 mSv and 8.8±5.4-times lower than for FDCT with 0.44±0.29 mSv (P<0.001) applying our institutional protocols.
CONCLUSIONS: Regarding image quality in postoperative assessment of CI, PCCT is noninferior to or in single features better than FDCT at a significantly lower radiation dose and shorter scan time.