利用阻抗测量引导人工耳蜗电极插入:一项计算仿真研究
Guiding Cochlear Implant Electrode Insertion Using Impedance Measurements: A Computational Simulation Study.
文献信息
| PMID | 43029671 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Enver Salkim |
| 作者单位 | Department of Electronic and Automation, Muş Alparslan University, 49100 Muş, Türkiye. |
| 期刊 | Sensors (Basel, Switzerland) |
| SCI 分区 | Q2 |
| IF | 4.1 |
| 研究类型 | 基础研究 · 基础/转化 |
| 所属专科 | 耳科 |
中文摘要
背景: 人工耳蜗(CI)插入过程中的术中组织创伤仍是残余听力损失的主要原因。由于插入目前主要依赖临床医生经验,且在光学上被遮挡的内耳内缺乏直接可视化,实时引导对于尽量减少与脆弱解剖结构的机械接触至关重要。本研究评估多通道电阻抗在自动电极插入过程中用于实时轨迹跟踪和边界接近检测的能力。方法:采用人耳蜗的计算有限元模型,分析八种双极配置(E1-E2至E15-E16)和25种插入模型(M1-M25)的复阻抗分布,在不同体积组织电导率下对中央、内侧和外侧轨迹进行2592次评估。结果:在1kHz时,电抗贡献可忽略不计(X<25 Ω,ϕ≤0.27°),证实以电阻性响应为主(R>2400 Ω,占总阻抗>99.5%)。体积电导率缩放基线阻抗幅度,但未显著改变空间分布几何形态。在三种插入轨迹中,阻抗呈现特征性U形演变,最初随着电极阵列(EA)向鼓阶(ST)中心推进而下降,随后在接近终止插入区域时升高。边界接近进一步增加阻抗变化,相对于中间基线,内侧轨迹的平均阻抗值增加24-38%,外侧轨迹增加25-35%。内侧对E3-E4(M20时65%)和外侧对E1-E2(M23时88%)观察到局部阻抗升高。结论:差异阻抗分布表明,边界诱导的接近效应显著超过基线空间变化,证明多通道阻抗测量对电极-壁接近的敏感性。这些发现支持将阻抗特征作为实时反馈信号用于边界接近检测、临床医生引导以及自动CI插入过程中闭环轨迹调整的潜在用途。识别特征性U形阻抗分布及其对边界接近的敏感性,可能为监测电极阵列位置和降低边界接触及相关机械创伤风险提供有价值的术中反馈。
英文摘要
Intraoperative tissue trauma during cochlear implant (CI) insertion remains a primary cause of residual hearing loss. Because insertion currently relies largely on clinician experience without direct visualization within the optically obscured inner ear, real-time guidance is critical for minimizing mechanical contact with delicate anatomical structures. This study evaluates multi-channel electrical impedance for real-time trajectory tracking and boundary proximity detection during automated electrode insertion. A computational finite-element model of the human cochlea was used to analyze complex impedance profiles across eight bipolar configurations (E1-E2 to E15-E16) and 25 insertion models (M1-M25), yielding 2592 evaluations across central, medial, and lateral trajectories under varying bulk tissue conductivities. At 1kHz, reactive contributions were negligible (X<25 Ω, ϕ≤0.27°), confirming an overwhelmingly resistive response (R>2400 Ω, >99.5% of total impedance). Bulk conductivity scaled baseline impedance magnitudes without substantially altering the spatial profile geometry. Across the three insertion trajectories, impedance exhibited a characteristic U-shaped evolution, initially decreasing as the electrode array (EA) advanced toward the scala tympani (ST) center before increasing near the terminal insertion region. Boundary proximity further increased impedance variation, with mean impedance values increasing by 24-38% for the medial trajectory and 25-35% for the lateral trajectory relative to the middle baseline. Localized impedance increases were observed for the medial pair E3-E4 (65% at M20) and the lateral pair E1-E2 (88% at M23). The differential impedance profiles indicate that boundary-induced proximity effects substantially exceed baseline spatial variations, demonstrating the sensitivity of multi-channel impedance measurements to electrode-wall proximity. These findings support the potential use of impedance signatures as a real-time feedback signal for boundary proximity detection, clinician guidance, and closed-loop trajectory adjustment during automated CI insertion. Recognizing the characteristic U-shaped impedance profile and its sensitivity to boundary proximity may provide valuable intraoperative feedback for monitoring electrode-array position and reducing the risk of boundary contact and associated mechanical trauma.