SmartNav引导人工耳蜗植入术中电极阵列插入动力学与电生理参数的术中相关性:一项探索性单中心研究

Intraoperative Correlates of Electrode Array Insertion Dynamics and Electrophysiological Parameters Under SmartNav-Guided Cochlear Implantation: An Exploratory Single-Centre Study.

临床研究耳科IF 3.6Q1

文献信息

中文摘要

背景/目的: 诸如SmartNav(Cochlear Ltd.,澳大利亚悉尼)之类的实时术中导航系统可在人工耳蜗植入过程中对电极阵列插入提供客观反馈,但目前尚不清楚其记录的参数是否会随术者所选择的装置层面和患者层面因素而变化。本研究比较了不同电极阵列设计(CI622细直型 versus CI632细蜗轴型)、手术入路(圆窗 versus 耳蜗开窗)以及患者年龄(儿童 versus 成人)之间的术中插入动力学和电生理参数。
方法: 分析了26例患者(3名儿童接受双侧植入)中连续29例在SmartNav监测下完成的植入。从设备日志和手术记录中提取插入角度、插入速度、总插入时间、阻抗在范围内的触点数量(22个中)以及可测得神经反应遥测(NRT)反应的触点数量(22个中);29例植入中有22例可获得插入动力学数据。数据分布为非正态,因此比较采用Mann-Whitney U检验并计算秩双列效应量、Fisher精确检验、Kruskal-Wallis检验和Spearman相关。此外还应用了探索性多变量回归、Brown-Forsythe方差检验、bootstrap效应量以及包含离群值和每例患者单耳敏感性分析;未进行多重比较校正。
结果: CI632阵列每电极推进速度快于CI622(中位数0.33 versus 0.56秒/电极;n=5 versus 16;p=0.015;秩双列r=-0.75;bootstrap 95% CI -1.00至-0.38),但当保留一个不合理的离群值时该差异不显著(p=0.113)。阻抗与NRT状态呈轻度相关(ρ=0.388,p=0.037,未校正),在校正年龄组后保持不变(偏ρ=0.389),但在每例患者单耳分析中不显著。本研究未能检测到年龄组之间或手术入路之间(5例耳蜗开窗 versus 24例圆窗插入)在任何参数上的差异,且阻抗状态在所有分层中变化极小。
结论: 在小型、不平衡队列的限制范围内,SmartNav衍生参数在患者年龄或手术入路之间未检测到差异。数据提示电极阵列设计与每电极插入节奏相关,但该观察依赖于一个被排除的数据点,需要进一步确认。阻抗-NRT关联为探索性,在考虑多重性后与偶然发现一致,目前尚不应被视为术中完整性信号。未将临床结局与术中参数相关联,因此无法就任一数组或入路的临床优越性作出推断。

英文摘要

Background/Objectives: Real-time intraoperative navigation systems such as SmartNav (Cochlear Ltd., Sydney, Australia) provide objective feedback on electrode array insertion during cochlear implantation, yet it remains unclear whether the parameters they record vary across the device- and patient-level factors a surgeon selects between. This study compared intraoperative insertion dynamics and electrophysiological parameters across electrode array design (CI622 slim-straight versus CI632 slim-modiolar), surgical approach (round window versus cochleostomy), and patient age (children versus adults). Methods: Twenty-nine consecutive implantations in 26 patients (three children implanted bilaterally) performed with SmartNav monitoring were analysed. Insertion angle, insertion speed, total insertion time, the number of contacts (of 22) with in-range impedance, and the number of contacts (of 22) with a measurable neural response telemetry (NRT) response were extracted from device logs and operative records; insertion kinetics were available for 22 of 29 implantations. Distributions were non-normal, so comparisons used the Mann-Whitney U test with rank-biserial effect sizes, Fisher's exact test, Kruskal-Wallis test, and Spearman correlation. Exploratory multivariable regression, Brown-Forsythe variance testing, bootstrap effect sizes, and outlier-inclusive and one-ear-per-patient sensitivity analyses were additionally applied; no correction for multiple comparisons was made. Results: The CI632 array was advanced faster per electrode than the CI622 (median 0.33 versus 0.56 s/electrode; n = 5 versus 16; p = 0.015; rank-biserial r = -0.75; bootstrap 95% CI -1.00 to -0.38), but the difference was not significant when one implausible outlying value was retained (p = 0.113). Impedance and NRT status were modestly correlated (ρ = 0.388, p = 0.037, uncorrected) and unchanged after adjustment for age band (partial ρ = 0.389), but were not significant in the one-ear-per-patient analysis. The study was unable to detect a difference between age groups or between surgical approaches (five cochleostomies versus 24 round window insertions) in any parameter, and impedance status varied minimally across all strata. Conclusions: Within the limits of a small, unbalanced cohort, SmartNav-derived parameters did not differ detectably across patient age or surgical approach. The data suggest that electrode array design is associated with per-electrode insertion pacing, an observation that depends on a single excluded data point and requires confirmation. The impedance-NRT association is exploratory, is consistent with a chance finding once multiplicity is considered, and should not yet be regarded as an intraoperative integrity signal. No clinical outcome was linked to the intraoperative parameters, so no inference about the clinical superiority of either array or approach can be drawn.