当前有源经皮骨传导植入物的安全性和有效性:一项系统综述。
Safety and efficacy of current active transcutaneous bone conduction implants: a systematic review.
文献信息
| PMID | 43026356 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Meghna Ray |
| 作者单位 | ENT Department, Sheffield Teaching Hospitals, (Royal Hallamshire Hospital) Glossop Road, Sheffield, S10 2TB, UK. |
| 期刊 | European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery |
| SCI 分区 | Q1 |
| IF | 2.3 |
| 研究类型 | 综述 Meta · 临床 |
| 所属专科 | 耳科 |
中文摘要
目的: 这是首个比较所有当前可用的有源经皮骨传导植入物(atBCI)(Bonebridge®、Osia®和Sentio®)并包含其中期安全性和有效性数据的系统综述。
方法: 检索了PubMed、EMBASE、Mendeley和Google Scholar,时间范围为2012年1月1日至2026年4月3日。遵循系统综述和荟萃分析首选报告项目(PRISMA)方案和Cochrane指南。纳入随机对照试验(RCT)、病例对照研究、队列研究和病例系列。文章由两名评价者筛选,并由资深作者监督。
结果: 在识别出的776篇论文中,去重后有66篇符合纳入标准并纳入最终研究。这些研究覆盖了1918例接受atBCI的患者。评估和结局指标以及检测时机存在显著异质性。53项研究报告了功能性听力增益。9项研究提供了足够的定量数据(包括变异度指标),可进行随机效应荟萃分析。合并的功能性听力增益为34.0 dB HL(95% CI 30.9-37.0 dB HL)。仅15项研究报告了言语接受阈的改善(范围为10.81至42)。并发症发生率非常低。
结论: 目前可用的atBCI似乎为其目标患者人群提供了可测量的听力康复获益。尽管所有当前可用的有源经皮骨传导植入系统均在定量合成中有所体现,但证据基础以Bonebridge研究为主。因此,该合并估计应被解释为对异质性证据总体的总体总结,而非各单个植入系统之间治疗效应等效的证据。
英文摘要
PURPOSE: This is the first systematic review comparing all currently available active transcutaneous bone conduction implants (atBCI) (Bonebridge®, Osia®, and Sentio®) with intermediate-term data on their safety and efficacy.
METHODS: PubMed, EMBASE, Mendeley, and Google Scholar were searched between 01 January 2012 and 03 April 2026. The Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) protocol and Cochrane guidance were followed. Included were randomised control trials (RCTs), case-control, cohort studies and case series. Articles were screened by two reviewers with senior author oversight.
RESULTS: Of the 776 papers identified, 66 met the inclusion criteria after deduplication and were included in the final study. The studies covered 1918 patients who had had atBCI. There was significant heterogeneity in assessment and outcome measures, as well as the timing of tests. Fifty-three studies reported functional hearing gain. Nine studies provided sufficient quantitative data, including measures of variance, to permit random-effects meta-analysis. The pooled functional hearing gain was 34.0 dB HL (95% CI 30.9-37.0 dB HL). Only 15 studies reported improvements in Speech Reception Threshold (ranging 10.81 to 42). Complications rates were very low.
CONCLUSION: The atBCIs available at present appear to provide measurable benefits in hearing rehabilitation for their targeted patient population. Although all currently available active transcutaneous bone conduction implant systems were represented in the quantitative synthesis, the evidence base was dominated by Bonebridge studies. Consequently, the pooled estimate should be interpreted as an overall summary across a heterogeneous body of evidence rather than as evidence of equivalent treatment effects across individual implant systems.