人工耳蜗磁铁保持问题的相关因素:单机构分析

Factors Associated With Magnet Retention Issues in Cochlear Implants: A Single Institution Analysis.

临床研究耳科IF 2.3Q2

文献信息

中文摘要

目的: 识别与接受人工耳蜗植入患者磁铁保持问题及高磁铁强度需求相关的临床和解剖因素。
方法: 对2018年至2023年间在一家三级学术医疗中心接受人工耳蜗植入的成年患者(n=168)进行回顾性病历审查。收集以下变量的数据:年龄、手术时BMI、术前头皮厚度、磁铁类型(轴向 vs. 径向)、处理器类型(耳背式[BTE] vs. 离耳式[OTE])以及术后保持问题(无法使用预期处理器、需要压迫/剃发、延迟激活或翻修手术)。统计分析通过卡方检验、Fisher精确检验和逻辑回归进行。
结果: 在168耳中,13.5%存在磁铁保持问题。48%的BTE使用者需要高磁铁强度(>3)或存在保持问题。OTE处理器比BTE有更高的保持问题率,尤其是使用径向磁铁时。年龄小于45岁(OR 6.88)和BMI≥30(OR 9.85)的患者磁铁保持问题显著更高。使用径向磁铁者风险远高于轴向磁铁者(17.6% vs. 2.0%),OTE处理器与保持失败强烈相关(23.1% vs. 4.3%)。
结论: 年龄<45岁和BMI≥30是高磁铁强度或保持问题的有用预测因素。临床医生可在术前咨询中使用这些参数,并考虑对这些高风险患者使用术前CT成像评估头皮厚度。对这些患者的医生咨询应包括保持问题增加的可能性,尤其是使用OTE处理器时。
证据等级: III。

英文摘要

OBJECTIVES: To identify clinical and anatomic factors associated with magnet retention issues and high magnet strength requirements in patients undergoing cochlear implantation.
METHODS: Retrospective chart review of adult patients (n = 168) who underwent cochlear implantation at a tertiary academic medical center between 2018 and 2023. Data was collected on the following variables: age, BMI at time of surgery, preoperative scalp thickness, magnet type (axial vs. diametric), processor type (behind-the-ear [BTE] vs. off-the-ear [OTE]), and postoperative retention issues (inability to use intended processor, need for compression/hair shaving, delayed activation, or revision surgery). Statistical analysis was performed via chi-square, Fisher's exact, and logistic regression.
RESULTS: Out of 168 ears, 13.5% had magnet retention issues. Forty-eight percent of BTE users needed a high magnet strength (> 3) or had retention issues. OTE processors had higher retention issues rates than BTE, especially with diametric magnets. Magnet retention was significantly higher for patients younger than 45 (OR 6.88) and with BMI ≥ 30 (OR 9.85). Those with diametric magnets were at much higher risk than those with axial magnets (17.6% vs. 2.0%) and OTE processors were strongly associated with retention failure (23.1% vs. 4.3%).
CONCLUSION: Age < 45 and BMI ≥ 30 are useful predictors of high magnet strength or retention issues. Clinicians may use these parameters during preoperative counseling and consider using preoperative CT imaging to assess scalp thickness for these high-risk patients. Physician counseling for these patients should include the likelihood of increased retention issues, especially with OTE processors.
LEVEL OF EVIDENCE: III.