气道狭窄改变常频及叠加喷射通气时潮气量、空气卷吸与氧源输送之间的平衡:一项台式模型研究
Airway stenosis alters the balance between tidal volume, air entrainment, and oxygen-source delivery during normal frequency and superimposed jet ventilation: a bench model study.
文献信息
| PMID | 43016537 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Mingyuan Yang |
| 作者单位 | Department of Anesthesiology, Emergency General Hospital, Beijing, China. |
| 期刊 | Frontiers in medicine |
| SCI 分区 | Q1 |
| IF | 3.7 |
| 研究类型 | 基础研究 · 基础/转化 |
| 所属专科 | 咽喉科 |
中文摘要
背景: 严重气道狭窄会改变开放系统喷射通气过程中的气体混合与容量输送。高频(HF)叠加及组件滴定如何改变这些效应仍未被充分描述。
目的: 比较单纯常频喷射通气(NFJV)与叠加高频喷射通气(SHFJV)在滴定常频组件(SHFJV-NF)或高频组件(SHFJV-HF)时,狭窄所致远端氧浓度(FiO2)、潮气量(VT)、压力、流量及估计气体组成的变化。
方法: 在开放硬质支气管镜台式模型中,分别在有和无90%固定气道狭窄条件下进行测试。单纯NFJV与SHFJV-NF共用56个常频参数设置;SHFJV-HF包括60个高频设置,常频组件固定。采用线性混合效应模型评估策略×气道交互作用及Holm校正的成对比较。另用匹配模型评估高频叠加。采用双源混合模型估计各设置下的气体组分分数和容量。
结果: 策略×气道交互作用对全部五项测量结局均显著(均P < 0.001)。在单纯NFJV、SHFJV-NF和SHFJV-HF中,狭窄所致FiO2变化中位数分别为+4.7、+29.1和+24.7个百分点,VT变化分别为-454、-577和-232 mL。在相同常频设置下,高频叠加使狭窄反应改变:FiO2 +24.1个百分点,VT -111.7 mL(95% CI -173.6至-49.8),实测呼气末正压(PEEP)+2.02 cmH2O,呼气流量幅度-8.11 L·min-1(均P < 0.001)。估计卷吸空气量在所有策略中均下降,而氧源容量并未一致增加。
结论: 在相同常频设置下,高频叠加放大了狭窄相关的FiO2升高、VT损失和PEEP升高。高频组件滴定在其测试参数范围内产生不同反应。估计的气体组成变化与氧稀释减少一致。远端FiO2不能证明通气或动脉氧合得以维持,且未测量CO2清除。
英文摘要
BACKGROUND: Severe airway stenosis alters gas mixing and volume delivery during open-system jet ventilation. How high-frequency (HF) superimposition and component titration modify these effects remains incompletely characterized.
OBJECTIVE: To compare stenosis-induced changes in distal oxygen concentration (FiO2), tidal volume (VT), pressure, flow, and estimated gas composition among pure normal-frequency jet ventilation (NFJV) and superimposed HF jet ventilation (SHFJV) with titration of either the NF (SHFJV-NF) or HF component (SHFJV-HF).
METHODS: An open rigid-bronchoscopy bench model was tested with and without 90% fixed airway stenosis. Pure NFJV and SHFJV-NF shared 56 NF parameter settings; SHFJV-HF included 60 HF settings with a fixed NF component. Linear mixed-effects models evaluated strategy × airway interactions and Holm-adjusted pairwise contrasts. A separate matched model assessed HF addition. Gas-component fractions and volumes were estimated at each setting using a two-source mixing model.
RESULTS: Strategy × airway interactions were significant for all five measured outcomes (all P < 0.001). Median stenosis-induced changes were +4.7, +29.1, and +24.7 percentage points for FiO2 and -454, -577, and -232 mL for VT during pure NFJV, SHFJV-NF, and SHFJV-HF, respectively. At identical NF settings, HF addition modified the stenosis response by +24.1 percentage points for FiO2, -111.7 mL for VT (95% CI -173.6 to -49.8), +2.02 cmH2O for measured positive end-expiratory pressure (PEEP), and -8.11 L·min-1 for expiratory flow magnitude (all P < 0.001). Estimated entrained-air volume decreased in all strategies, whereas oxygen-source volume did not increase uniformly.
CONCLUSION: At identical NF settings, HF addition amplified the stenosis-associated FiO2 rise, VT loss, and PEEP increase. HF-component titration produced a different response across its tested parameter range. Estimated gas-composition changes were consistent with reduced oxygen dilution. Distal FiO2 did not establish preserved ventilation or arterial oxygenation, and CO2 elimination was not measured.