人工耳蜗植入者中的多音电耳蜗描记术:专家评估与基于算法的比较

Multitone Electrocochleography in Cochlear Implant Recipients: An Expert and Algorithm-Based Comparison.

临床研究耳科IF 1.9Q2

文献信息

中文摘要

目的: 本研究比较了人工耳蜗植入者术后纯音与多音电耳蜗描记术(ECochG)记录在信号特征及频率特异性反应可检测性方面的差异。评估了专家视觉评估与基于傅里叶变换的分箱算法在检测ECochG反应及识别多音记录中频谱成分方面的可靠性。
研究设计: 前瞻性初步队列研究。
地点: 三级转诊医院。
患者: 21例植入后至少6个月残余听力稳定的人工耳蜗植入者。
干预: 通过植入电极阵列记录耳蜗内耳蜗微音电位,使用250、500、1000和2000 Hz纯音以及包含上述频率中3个或全部4个的2种多音刺激。测量在植入后至少6个月进行,并在其后2至48小时及2至4个月随访。
主要结局指标: 比较专家与算法对反应的分类。评估评分者间一致性、信噪比、敏感性和特异性。
结果: 在所有参与者中,所有测试频率均检测到纯音ECochG反应。多音刺激在21例参与者中的17例产生了总体ECochG反应。反应幅度和空间模式随电极位置和残余听力而变化。虽然观察到音调定位趋势,但最大反应常向基底耳蜗偏移。纯音的评分者间一致性较高,但多音记录中频率特异性成分的一致性明显较低,尤其是在低频。所有条件下的信噪比均较低。分箱算法对纯音显示出高敏感性但特异性有限,对多音刺激则呈现频率依赖性的权衡。
结论: 在人工耳蜗植入至少6个月后听力图和电极位置稳定的患者中,多音ECochG可行并能产生可检测的反应。然而,信号特征随电极位置和残余听力而变化,产生比纯音更复杂的模式。这种复杂性阻碍了专家和客观方法对单个频谱成分的可靠识别。

英文摘要

OBJECTIVES: This study compared postoperative pure-tone and multitone electrocochleography (ECochG) recordings in cochlear implant recipients regarding signal characteristics and detectability of frequency-specific responses. The reliability of expert visual assessment and a Fourier transform-based binning algorithm were evaluated for detecting ECochG responses and identifying spectral components in multitone recordings.
STUDY DESIGN: Prospective pilot cohort study.
SETTING: Tertiary referral hospital.
PATIENTS: Twenty-one cochlear implant recipients with stable residual hearing at least 6 months postimplantation.
INTERVENTIONS: Intracochlear cochlear microphonics potentials were recorded through the implant electrode array using pure tones at 250, 500, 1000, and 2000 Hz and 2 multitone stimuli comprising 3 or all 4 of the aforementioned frequencies. Measurements were conducted at least 6 months postimplantation, with follow-ups at 2 to 48 hours and 2 to 4 months thereafter.
MAIN OUTCOME MEASURES: Response classification by the expert and the algorithm was compared. Inter-rater agreement, signal-to-noise ratio, sensitivity, and specificity were assessed.
RESULTS: Pure-tone ECochG responses were detected at all tested frequencies in every participant. Multitone stimulation produced an overall ECochG response in 17 of 21 participants. Response amplitudes and spatial patterns varied with electrode position and residual hearing. While tonotopic trends were observed, maximal responses were frequently shifted toward the basal cochlea. Inter-rater agreement was high for pure tones but clearly lower for frequency-specific components of multitone recordings, particularly at low frequencies. Signal-to-noise ratios were low across all conditions. The binning algorithm showed high sensitivity but limited specificity for pure tones and a frequency-dependent trade-off for multitone stimulation.
CONCLUSIONS: In patients with stable audiometry and electrode position after at least 6 months of cochlear implant insertion, multitone ECochG was feasible and produced detectable responses. However, signal characteristics varied with electrode position and residual hearing, yielding more complex patterns than pure tones. This complexity hindered reliable identification of individual spectral components by experts and by objective methods.