防止电极阵列脱出和旋转的人工耳蜗电极固定新技术

New Technique for Fixation of Cochlear Implant Electrodes Preventing Extrusion and Rotation of the Array.

临床研究耳科IF 1.3Q3

文献信息

中文摘要

背景: 人工耳蜗(CI)已发展为应对多种听力障碍,适应证扩展至包括有残余听力的患者。保留残余听力可实现电声刺激,可能改善CI效果。虽然“软手术”技术被广泛采用以尽量减少即刻耳蜗创伤,但在乳突内盘绕电极导线过程中插入后电极移动仍研究不足。本研究提出新的电极固定手术技术,以尽量减少插入后移动。
方法: 一项与Cochlear®(比利时梅赫伦)合作的初步研究,检验了在盘绕过程中于乳突内设置稳定沟槽以及在面后鼓室切开处使用可移除稳定材料的可行性。目标包括尽量减少预弯型蜗轴周围532/632电极的轴向旋转、改善电极导线固定、防止术中插入深度变化,并评估儿童适用性。
结果: 两种拟议技术的联合应用实现了稳定的电极放置。在面后鼓室切开处使用可移除无菌海绵可防止电极轴向旋转超过90°以及插入深度改变,尤其是在将电极放置于上鼓室沟槽以及随后在乳突内盘绕多余导线期间。该研究还证实了这些技术在儿童模型中的可行性。
结论: 所提出的技术似乎能有效确保插入后电极在耳蜗内的放置稳定。这些技术仅需对标准手术流程做微小调整,无需昂贵材料或显著延长手术时间。因此作者认为其具有系统性实施的潜力。仍需进一步研究以验证这些技术并探索其对整体CI性能的影响。

英文摘要

BACKGROUND: Cochlear implants (CIs) have evolved to address diverse hearing impairments, expanding indications to include patients with residual hearing. Preserving residual hearing enables electro-acoustic stimulation, potentially improving CI outcomes. While "soft surgery" techniques are widely adopted to minimize immediate cochlear trauma, post-insertion electrode movements during the coiling of the electrode lead in the mastoid remain understudied. This study proposes novel surgical techniques of electrode fixation to minimize post-insertion movement.
METHODS: A pilot study, in collaboration with Cochlear® (Mechelen, Belgium), examined the feasibility of stabilization grooves in the mastoid and removable stabilization material in the posterior tympanotomy during coiling. Objectives included minimizing axial rotation of the pre-curved perimodiolar 532/632 electrodes, improving electrode lead fixation, preventing intraoperative changes in the insertion depth, and assessing pediatric applicability.
RESULTS: Combination of both proposed techniques allowed for stable electrode placement. The use of a removable sterile sponge in the posterior tympanotomy prevented axial rotation of the electrode beyond 90° and alterations in insertion depth, particularly during placement of the electrode in the attical groove and the subsequent coiling of the excess lead in the mastoid. The study also confirmed the feasibility of these techniques in pediatric models.
CONCLUSION: The proposed techniques seem to be effective in securing electrode placement in the cochlea after insertion. These techniques involve only minor adaptations to the standard procedure, without the need for costly material or a substantially longer operating time. The authors therefore believe they have the potential for systematic implementation. Further research is required to validate these techniques and explore their impact on overall CI performance.