Assessing the feasibility of uTPS for volumetric modulated arc therapy treatment plan optimization of dual-layer multileaf collimator linear accelerators.
文献信息
| PMID | 43004445 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Qingxin Wang |
| 作者单位 | Department of Radiation Oncology, Tianjin Medical University Cancer Institute & Hospital, National Clinical Research Center for Cancer, Tianjin's Clinical Research Center for Cancer, Key Laboratory of Cancer Prevention and Therapy, Tianjin, 300060, China. |
| 期刊 | Radiation oncology (London, England) |
| SCI 分区 | Q1 |
| IF | 3.9 |
| 研究类型 | 临床研究 · 临床 |
| 所属专科 | 咽喉科 |
中文摘要
背景: 双层多叶准直器(MLC)加速器(如Halcyon)在临床中的应用日益增多;然而,治疗计划系统(TPS)能否有效优化此类加速器的治疗计划仍是一个重要考量。
目的: 本研究评估使用uTPS优化双层MLC加速器治疗计划的可行性,并以Eclipse作为临床参考。
方法: 使用Halcyon 2.0加速器的标准放射数据在uTPS(上海联影医疗科技股份有限公司,版本R001)中建模射束。选取20例病例,其中下咽癌、食管癌、乳腺癌和宫颈癌各5例。使用uTPS随机平台优化器和Eclipse光子优化器(PO)生成容积旋转调强放疗计划。所有计划均使用Acuros外照射(AXB)算法进行计算以进行剂量学比较。分析并比较关键指标,如适形指数(CI)、均匀性指数(HI)和危及器官(OAR)剂量。验证了在Halcyon加速器上执行uTPS计划的可行性。使用ArcCheck确认剂量学准确性,并进行Wilcoxon符号秩检验及Benjamini-Hochberg错误发现率校正。
结果: 对于食管癌和宫颈癌病例,uTPS与Eclipse计划表现出相似的剂量学质量,CI、HI和OAR剂量无显著差异。然而,uTPS在特定病例中表现出更优的OAR保护:在下咽癌中,uTPS显著降低脑干[公式:见正文] 34.48%、左侧腮腺[公式:见正文] 11.8%和左侧腮腺[公式:见正文] 6.9%(均P < 0.05);在乳腺癌中,uTPS显著降低脊髓[公式:见正文] 49.6%(P < 0.05)。其他剂量学参数未观察到统计学显著差异。uTPS与Eclipse计划之间的总机器跳数未观察到显著差异。所有uTPS计划均在Halcyon加速器上成功执行,伽马通过率超过97%。
结论: 市售uTPS支持双层MLC Halcyon加速器的有效计划优化,其计划质量与Eclipse相当。
英文摘要
BACKGROUND: Dual-layer multileaf collimator (MLC) accelerators, such as Halcyon, are increasingly used in clinical settings; however, the ability of treatment planning systems (TPS) to effectively optimize treatment plans for such accelerators remains an important consideration.
PURPOSE: This study evaluates the feasibility of using uTPS to optimize treatment plans for dual-layer MLC accelerators, with Eclipse as a clinical reference.
METHODS: Standard radiation data from the Halcyon 2.0 accelerator were used to model the beam in the uTPS (Shanghai United Imaging Healthcare Co., Ltd., version R001). Twenty cases were selected, with five each for hypopharyngeal, esophageal, breast, and cervical cancers. Volumetric modulated arc therapy plans were generated using the uTPS stochastic platform optimizer and the Eclipse photon optimizer (PO). All plans were calculated using the Acuros external beam (AXB) algorithm for dosimetric comparison. Key indicators, such as the conformity index (CI), homogeneity index (HI), and organ-at-risk (OAR) doses, were analyzed and compared. The feasibility of delivering uTPS plans on the Halcyon accelerator was validated. The dosimetric accuracy was confirmed using ArcCheck, and Wilcoxon signed-rank tests with Benjamini-Hochberg false discovery rate correction were performed.
RESULTS: uTPS and Eclipse plans exhibited similar dosimetric qualities for esophageal and cervical cancer cases, with no significant differences in CI, HI, and OAR doses. However, uTPS demonstrated superior OAR sparing in specific cases: in hypopharyngeal cancer, uTPS significantly reduced the brainstem [Formula: see text] by 34.48%, the left parotid[Formula: see text] by 11.8%, and the left parotid [Formula: see text] by 6.9% (all P < 0.05); in breast cancer, uTPS significantly reduced the spinal cord [Formula: see text]by 49.6% (P < 0.05). No statistically significant differences were observed in other dosimetric parameters. No significant differences in total monitor units were observed between the uTPS and Eclipse plans. All uTPS plans were successfully delivered on the Halcyon accelerator, with gamma passing rates exceeding 97%.
CONCLUSIONS: The commercially available uTPS supports effective plan optimization for dual-layer MLC Halcyon accelerators, offering plan quality comparable to that of Eclipse.