比较用于术中监测浮动质量传感器耦合效能的声处理器:基于听性脑干反应

Comparison of sound processors for intraoperative monitoring of floating mass transducer coupling efficacy using auditory brainstem responses.

临床研究耳科IF 3.4Q1

文献信息

中文摘要

目的: Vibrant Soundbridge(VSB)有源中耳植入物的获益取决于浮动质量传感器与中耳连接点之间的耦合质量。已提出通过听性脑干反应进行术中评估,可使用已编程的音频处理器或专用研究工具(AcoustiAP)。由于最初使用的处理器(AP 404)已停产,本研究比较两种后续设备(SAMBA 2 Hi、SAMBA 2 Lo)和 AcoustiAP 相对于 AP 404 的传输特性。
方法: 将四种设备(MED-EL,Innsbruck)耦合至 Aurical HIT 测试箱的 2-cc 耦合腔。以 0、30 和 65 dB nHL 呈现 CE-Chirp 刺激;还记录了 AcoustiAP 的 0-100 dB nHL 输入-输出函数。使用 DSLv5 对声处理器进行编程,阈值设为 25 dB HL。根据 DIN EN 60118-7 计算高频平均值(HFA)。
结果: 固有噪声 HFA 分别为 12.0、22.5、25.7 和 26.0 dB SPL(SAMBA 2 Lo、AP 404、AcoustiAP、SAMBA 2 Hi)。在 30 dB nHL 时,各声处理器产生相当输出(46.5-47.0 dB SPL),而 AcoustiAP 仍处于其噪声底限内。SAMBA 2 Hi 表现出最线性的行为;SAMBA 2 Lo 表现出明显压缩。AcoustiAP 信号在约 35 dB nHL 以下被噪声掩盖,并在 70.1 dB SPL 饱和。
结论: SAMBA 2 Hi 是 AP 404 的合适替代品。由于 SAMBA 2 Lo 的设计,其输出水平和噪声底限低于其 Hi 对应型号。这可能为根据听力损失程度测试个体化编程处理器提供选择。这一想法应在进一步研究中探讨。AcoustiAP 目前尚无法在无校正的情况下实现可靠的基于阈值的评估。

英文摘要

OBJECTIVE: The benefit of the Vibrant Soundbridge (VSB) active middle ear implant depends on the coupling quality between the floating mass transducer and the connection point in the middle ear. Intraoperative assessment via auditory brainstem responses has been proposed using either a programmed audio processor or a dedicated research tool (AcoustiAP). As the originally used processor (AP 404) is discontinued, this study compares the transmission characteristics of two successor devices (SAMBA 2 Hi, SAMBA 2 Lo) and the AcoustiAP against the AP 404.
METHODS: The four devices (MED-EL, Innsbruck) were coupled to a 2-cc coupler of an Aurical HIT test box. CE-Chirp stimuli were presented at 0, 30 and 65 dB nHL; a 0-100 dB nHL input-output function was also recorded for the AcoustiAP. The sound processors were programmed using DSLv5 for a threshold of 25 dB HL. High-frequency averages (HFA) were calculated according to DIN EN 60118-7.
RESULTS: Intrinsic noise HFA were 12.0, 22.5, 25.7 and 26.0 dB SPL (SAMBA 2 Lo, AP 404, AcoustiAP, SAMBA 2 Hi). At 30 dB nHL, the sound processoes produced comparable outputs (46.5-47.0 dB SPL), whereas the AcoustiAP remained within its noise floor. The SAMBA 2 Hi showed the most linear behaviour; the SAMBA 2 Lo showed pronounced compression. The AcoustiAP signal was embedded in noise up to about 35 dB nHL and saturated at 70.1 dB SPL.
CONCLUSIONS: The SAMBA 2 Hi is a suitable replacement for the AP 404. Due to the design of the SAMBA 2 Lo its output level and noise floor are lower compared to its Hi counter variant. That potentially offers options for testing individual programmed processors based on the degree of hearing loss. This idea should be investigated in further studies. The AcoustiAP currently does not allow reliable threshold-based assessment without correction.