南印度与苏格兰人群耳蜗管长度的计算机断层扫描比较分析:对人工耳蜗电极选择的意义
Comparative analysis of cochlear duct length using computed tomography in South Indian and Scottish populations: Implications for cochlear implant electrode selection.
文献信息
| PMID | 43000557 |
|---|---|
| 原文 | 在 PubMed 查看原文 ↗ |
| 发表日期 | 2026 |
| 作者 | Aparajith Sathish Kumar |
| 作者单位 | School of Medicine, University of Dundee, Dundee, Scotland. 2489335@dundee.ac.uk. |
| 期刊 | European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery |
| SCI 分区 | Q1 |
| IF | 2.3 |
| 研究类型 | 临床研究 · 临床 |
| 所属专科 | 耳科 |
中文摘要
目的: 耳蜗管长度(CDL)在个体和人群间存在差异,影响人工耳蜗(CI)电极选择、插入深度及术后结局。大多数CI电极阵列基于白种人解剖数据设计,来自南亚人群的证据有限。本研究旨在比较南印度与苏格兰队列的CDL,并探讨对人群特异性CI电极选择的意义。
方法: 一项回顾性队列研究分析了颞骨CT扫描,包括南印度队列的100个耳蜗和苏格兰队列的100个耳蜗。CDL采用Alexiades公式估算,该公式基于沿蜗轴中轴在冠状位CT图像上测量的A值。所有测量均由资深顾问独立验证,以确保准确性和可重复性。
结果: 南印度组平均CDL(26.38 ± 2.00 mm)显著短于苏格兰组(33.07 ± 1.19 mm),绝对差异为6.70 mm(p < 0.001)。南印度队列相应的A值也较小(7.30 ± 0.48 mm vs. 8.91 ± 0.28 mm;p < 0.001)。
结论: 南印度与苏格兰队列之间估算的CDL存在显著差异。这些形态学发现表明所研究人群之间存在解剖变异。然而,其对人工耳蜗电极选择和临床结局的意义尚需前瞻性验证。
英文摘要
PURPOSE: Cochlear duct length (CDL) varies among individuals and populations, influencing cochlear implant (CI) electrode choice, insertion depth, and postoperative outcomes. Most CI electrode arrays are designed using Caucasian anatomical data, with limited evidence from South Asian populations. This study aimed to compare CDL between South Indian and Scottish cohorts and explore implications for population-specific CI electrode selection.
METHODS: A retrospective cohort study analysed temporal bone CT scans comprising 100 cochleae from the South india cohort and 100 cochleae from the Scottish cohort. CDL was estimated using the Alexiades formula, derived from A-values measured on coronal CT images along the mid-modiolar axis. All measurements were validated independently by senior consultants to ensure accuracy and reproducibility.
RESULTS: Mean CDL was significantly shorter in the South Indian group (26.38 ± 2.00 mm) compared with the Scottish group (33.07 ± 1.19 mm), showing an absolute difference of 6.70 mm (p < 0.001). The corresponding A-values were also smaller in the South Indian cohort (7.30 ± 0.48 mm vs. 8.91 ± 0.28 mm; p < 0.001).
CONCLUSION: Significant differences in estimated CDL were identified between the South Indian and Scottish cohorts. These morphometric findings demonstrate anatomical variation between the studied populations. However, their implications for cochlear implant electrode selection and clinical outcomes require prospective validation.