基底最末端电极位置对人工耳蜗植入中角插入深度预测准确性的影响

Influence of basal-most electrode positioning on the accuracy of angular insertion depth predictions in cochlear implantation.

临床研究耳科IF 2.3Q1

文献信息

中文摘要

目的: 准确预测角插入深度(AID)对于优化人工耳蜗(CI)植入效果至关重要。然而,术前预测与术后结果之间仍存在差异。在本研究中,我们系统评估了基于公式的AID预测的准确性,以确定预测误差的主要来源。
方法: 在这项回顾性研究中,使用基于公式的方法生成各种MED-EL电极阵列的术前AID预测,假设电极止动器位于圆窗处并完全插入。将这些预测结果与通过数字容积断层扫描获得的术后实测AID进行比较。研究的主要变量是最靠近基底位置的电极的线性插入深度(LIDE12),定义为其距圆窗的距离。
结果: LIDE12范围为-13.8 mm至11.0 mm;负值表示不完全插入,在9%的病例中观察到这种情况。预测AID与实测AID之间的中位差异为-11°(四分位距74°)。在通过纳入实测LIDE12而非假设插入至止动器来校正预测后,中位差异变为+35°,四分位距显著缩小(37°)。
结论: 虽然当前的AID预测模型总体上显示出临床可接受的准确性,但仍可能出现相当大的差异。一个主要影响因素是电极阵列的实际手术放置。这种由LIDE12测量所反映的变异性表明,电极放置的实际结果往往与假设的放置不同,在规划电极位置时应考虑这一因素。

英文摘要

OBJECTIVE: Accurate prediction of the angular insertion depth (AID) is essential for optimising cochlear implant (CI) outcomes. However, discrepancies between preoperative predictions and postoperative results persist. In this study, we evaluated systematically the accuracy of equation-based AID prediction to identify the main contributors to prediction errors.
METHODS: In this retrospective study, preoperative AID predictions for various MED-EL electrode arrays were generated by using an equation-based method, assuming full insertion with the electrode stopper at the round window. These predictions were compared with postoperatively measured AIDs obtained by digital volume tomography. The primary variable investigated was the linear insertion depth of the electrode closest to the basal position (LIDE12), defined as its distance from the round window.
RESULTS: The LIDE12 ranged from - 13.8 mm to 11.0 mm; negative values indicate incomplete insertions, which were observed in 9% of cases. Between predicted and measured AID a median difference of - 11° (interquartile range 74°) was observed. After correcting the prediction by incorporating the measured LIDE12 instead of assuming insertion to the stopper, the median difference shifted to + 35° with a markedly reduced interquartile range (37°).
CONCLUSION: While current AID prediction models generally show clinically acceptable accuracy, considerable discrepancies can still occur. One major contributing factor is the actual surgical placement of the electrode array. This variability, captured by the LIDE12 measurement, demonstrates that the real-life result of the electrode placement often differs from the assumed placing, and that this factor should be considered when planning electrode positioning.