同期人工耳蜗植入与前庭神经鞘瘤切除术患者的进阶编程需求

Advanced programming requirements of subjects who undergo simultaneous cochlear implantation and vestibular schwannoma resection.

临床研究耳科IF 1.8Q3

文献信息

中文摘要

引言: 研究前庭神经鞘瘤切除术(VSR)后立即行人工耳蜗植入(CI)的研究显示,言语理解改善前景可观但往往存在差异。目前尚无已知研究探讨该患者群体的编程需求。我们报告了为在VSR之前接受CI的患者提供声音所需的进阶CI编程。
方法: 本描述性研究分析了12例单侧前庭神经鞘瘤接受同期CI和VSR的患者。对12例中8例通过CI获得声音的患者分析了编程参数和结局。编程参数包括活动电极数量、脉宽、刺激速率、最大值以及达到较低(阈值)和较高刺激水平(C水平)所需的电荷水平。患者结局包括CI声场阈值、术后CNC Words和AzBio Sentences得分、数据记录以及言语、空间与听觉质量量表(SSQ)问卷得分。
结果: 所有8例通过CI获得声音的患者均需要一定程度的进阶编程,包括增宽脉宽、停用电极以及降低最大值和刺激速率。在VSR之前接受CI的患者的刺激水平显著高于文献中报告的非肿瘤患者。
结论: 在VSR之前接受CI的患者存在多种编程挑战,并表现出可变的术后言语得分。所有8例通过CI获得声音的患者均从该手术中获益,这由设备使用、结局和SSQ-12反应所证明。

英文摘要

INTRODUCTION: Studies examining cochlear implantation (CI) immediately following vestibular schwannoma resection (VSR) demonstrate promising yet often variable improvements in speech understanding. There are no known existing studies that address the programming requirements for this patient population. We report on the advanced CI programming required to provide sound for patients who underwent CI prior to VSR.
METHODS: This descriptive study analyzed 12 patients with unilateral vestibular schwannoma who underwent simultaneous CI and VSR. Programming parameters and outcomes were analyzed in 8 of 12 patients who had sound with the CI. Programming parameters included number of active electrodes, pulse width, stimulation rate, maxima, and charge levels required for lower (threshold) and upper stimulation levels (C levels). Patient outcomes included CI sound field thresholds, post-operative scores obtained on CNC Words and AzBio Sentences, datalogging, and scores for the Speech, Spatial and Qualities of Hearing Scale (SSQ) questionnaire.
RESULTS: All 8 patients who received sound with their CI required some level of advanced programming, including widening of pulse width, deactivation of electrodes, and reduction of maxima and stimulation rate. Stimulation levels for patients who received a CI prior to VSR were significantly higher than those reported in the literature for non-tumor patients.
CONCLUSION: Patients who received a CI prior to VSR present with various programming challenges and demonstrate variable post-operative speech scores. All 8 patients who received sound with the CI benefited from this procedure, as demonstrated by device use, outcomes, and SSQ-12 responses.