中文摘要
目的: 本研究提出一种基于解剖的拟合精细方法(rfABF),旨在减少双耳电极阵列植入不对称(两耳间常与耳蜗覆盖不完全相关)的双侧人工耳蜗(CI)使用者的双耳位置失配。
设计: 前瞻性单中心概念验证研究,比较rfABF与常规临床拟合(CBF)。ABF映射采用两种算法推导方法制备:一种缩放电极间频谱间距(zipped),另一种转置临床频率与音调拓扑频率之间的比值(transposed)。随后根据使用者反馈对这些映射进行精细调整,直至获得自然的音质。评估包括安静与噪声下的言语感知、声源定位、主观音质评分(HISQUI19),以及使用瞳孔测量法评估的聆听努力。
研究样本: 九名经验丰富的成年双侧CI使用者,其电极阵列植入深度和/或电极阵列类型存在双耳不对称。
结果: rfABF使双侧言语理解在安静条件下显著改善(CBF:77.5% ± 8.1%;rfABF:83.1% ± 8.0%;p = 0.025),并在S0N0(信号与噪声均来自前方)和S0Ncontra(信号来自前方、噪声来自对侧耳)两种噪声配置下均显著改善。rfABF使双耳刺激位置失配从平均33.8° ± 27.3°大幅降至6.4° ± 4.2°。探索性、描述性瞳孔测量表明,部分参与者的聆听努力降低,表现为事件相关瞳孔扩张(ERPD)波幅降低,尤其是在言语可懂度得以保持或改善的病例中。
结论: rfABF显示出临床潜力,可改善电极阵列植入不对称的双侧CI使用者的言语理解并促进其获取双耳线索,并有初步证据表明可降低聆听努力。
英文摘要
OBJECTIVE: This study proposes a refined approach to anatomy-based fitting (rfABF) that aims to reduce interaural place mismatch in bilateral cochlear implant (CI) users with asymmetric electrode array insertions between ears, a condition frequently associated with incomplete cochlear coverage.
DESIGN: Prospective monocenter proof-of-concept study comparing rfABF with conventional clinically based fitting (CBF). ABF maps were prepared using two algorithmically derived methods: one scaling inter-electrode spectral spacing (zipped), and another transposing the ratio between clinical and tonotopic frequencies (transposed). These maps were then refined based on user feedback until a natural sound quality was achieved. Assessments consisted of speech perception in quiet and noise, sound localization, subjective sound quality ratings (HISQUI19), and listening effort assessed using pupillometry.
STUDY SAMPLE: Nine experienced adult bilateral CI users with interaural asymmetry in electrode array insertion depth and/or electrode array type.
RESULTS: Bilateral speech understanding improved significantly with rfABF in quiet (CBF: 77.5% ± 8.1%; rfABF: 83.1% ± 8.0%; p = 0.025) and in noise under both S0N0 (signal and noise from front) and S0Ncontra (signal from front, noise from contralateral ear) configurations. Interaural place-of-stimulation mismatch was substantially reduced from a mean of 33.8° ± 27.3° to 6.4° ± 4.2° with rfABF. Exploratory, descriptive pupillometry indicated reduced listening effort in a subset of participants, reflected by lower event-related pupil dilation (ERPD) amplitudes, particularly in cases where speech intelligibility was preserved or improved.
CONCLUSIONS: rfABF shows clinical potential to improve speech understanding and facilitate access to binaural cues in bilateral CI users with asymmetric electrode array insertions, with preliminary evidence for reduced listening effort.