使用光子计数探测器CT改善对位置异常的中耳假体的可视化:一项尸体研究。

Improved visualization of malpositioned middle ear prostheses using photon-counting detector CT: a cadaver study.

临床研究耳科IF 1.8Q3

文献信息

中文摘要

背景: 中耳假体的评估需要高分辨率成像来评估位置和结构完整性。
目的: 在尸体标本中评估光子计数探测器(PCD)-CT对中耳假体成像的诊断性能。
材料与方法: 将9种不同的中耳假体植入5个新鲜冷冻全头标本中,分别置于正确的解剖位置和移位以模拟位置异常。标本使用PCD-CT(NAEOTOM Alpha,Siemens)和能量积分探测器(EID)-CT(SOMATOM Force,Siemens)扫描,分别重建为0.2 mm和0.4 mm。每台扫描仪评估18个病例(9个假体,2种位置),由2名神经放射科医生评估(36次判读),他们对扫描仪和位置设盲,并对假体位置、判读者信心和整体图像质量进行评分。比较了与参考标准的一致性以及判读者之间的一致性。
结果: 与EID-CT相比,PCD-CT显示出更高的诊断信心,而准确性和判读者间一致性在扫描仪之间未显示出显著差异。正确判读,PCD-CT与EID-CT:外侧接触,31/36(86.1%;95% CI,58.3-94.4)对29/36(80.6%;95% CI,55.6-91.7),P = .500;内侧接触,30/36(83.3%;95% CI,55.6-94.4)对26/36(72.2%;95% CI,50.0-83.3),P = .250。PCD-CT识别出所有完全接触丧失(评分3)的病例,而EID-CT每次接触漏诊1次判读。PCD-CT的判读者间一致性可能高于EID-CT(外侧:κw:0.804 [95% CI,0.578-0.941] 对0.571 [95% CI,0.302-0.778],P = .250;内侧:0.684 [95% CI,0.191-0.934] 对0.481 [95% CI,0.024-0.835],P = .250)。PCD-CT的判读者信心显著更高;判读者1和2的平均配对差异分别为外侧接触9.4(95% CI,1.8-17.1;Holm校正P = .019)和23.1分(95% CI,13.4-32.7;P < .001),内侧接触16.4(95% CI,10.0-22.8;P < .001)和17.8分(95% CI,6.8-28.8;P = .007)。判读者分别在16/18(88.9%;95% CI,65.3-98.6)和17/18例(94.4%;95% CI,72.7-99.9)中偏好PCD-CT图像质量。
结论: PCD-CT与显著更高的判读者信心和更偏好的图像质量相关,而在诊断准确性或判读者间一致性方面无显著差异。这些发现表明PCD-CT是术后中耳评估的一种有前景的模态,值得进一步研究。

英文摘要

BACKGROUND: Evaluation of middle ear prostheses requires high-resolution imaging to assess position and structural integrity.
PURPOSE: To assess the diagnostic performance of photon-counting-detector (PCD)-CT for middle ear prosthesis imaging in cadaveric specimens.
MATERIALS AND METHODS: Nine different middle ear prostheses were implanted in 5 fresh-frozen whole-head specimens, both in correct anatomic position and displaced to mimic malposition. Specimens were scanned on PCD-CT (NAEOTOM Alpha, Siemens) and energy-integrating-detector (EID)-CT (SOMATOM Force, Siemens), reconstructed at 0.2 mm and 0.4 mm, respectively. Eighteen cases (9 prostheses, 2 positions) per scanner were evaluated by 2 neuroradiologists (36 readings), blinded to scanner and position, who scored prosthesis position, reader confidence, and overall image quality. Agreement with the reference standard and between readers was compared.
RESULTS: PCD-CT showed higher diagnostic confidence compared with EID-CT, while accuracy and inter-reader agreement did not demonstrate significant differences between scanners. Correct readings, PCD-CT versus EID-CT: lateral contact, 31/36 (86.1%; 95% CI, 58.3-94.4) vs 29/36 (80.6%; 95% CI, 55.6-91.7), P = .500; medial contact, 30/36 (83.3%; 95% CI, 55.6-94.4) vs 26/36 (72.2%; 95% CI, 50.0-83.3), P = .250. PCD-CT identified all cases of complete contact loss (score 3), while EID-CT missed 1 reading per contact. Inter-reader agreement may be higher with PCD-CT vs EID-CT (lateral: κw: 0.804 [95% CI, 0.578-0.941] vs 0.571 [95% CI, 0.302-0.778], P = .250; medial: 0.684 [95% CI, 0.191-0.934] vs 0.481 [95% CI, 0.024-0.835], P = .250). Reader confidence was significantly higher with PCD-CT; mean paired differences for Readers 1 and 2 were 9.4 (95% CI, 1.8-17.1; Holm-adjusted P = .019) and 23.1 points (95% CI, 13.4-32.7; P < .001) for lateral contact, and 16.4 (95% CI, 10.0-22.8; P < .001) and 17.8 points (95% CI, 6.8-28.8; P = .007) for medial contact. Readers preferred PCD-CT image quality in 16/18 (88.9%; 95% CI, 65.3-98.6) and 17/18 cases (94.4%; 95% CI, 72.7-99.9), respectively.
CONCLUSION: PCD-CT was associated with significantly higher reader confidence and preferred image quality, without significant differences in diagnostic accuracy or inter-reader agreement. These findings suggest that PCD-CT is a promising modality for postoperative middle ear evaluations and warrants further investigations.